800: 5mm TXI-Z, #550501-0011
In principle: Do NOT trust these values blindly - verify them if questionable!
Amplifiers can go bad or your sample environment like salt, temperature, ... might alter the pulse width!
1. General Information:
- Shimfile: 550501_011ws.shim is a good starting point for aqueous samples
2. Hard Pulses:
Nucleus | Type | Attenuation | Channel | 90deg Pulse | Amplifier | Calculated | Date |
---|
1H | | -1.0 dB | f1 | 9.4 us | 100W | | 11/23/05 |
| waltz16 | | f1 | 70.0 us | 100W | | |
| TOCSY | | f1 | 25-35.0 us | 100W | | |
| ROESY | | f1 | 100.0 us | 100W | | |
| | | | | | | . |
13C | hard | -3.3 dB | f2 | 15.0 us | 500W | | 11/23/05 |
| Spec | -5.2 dB | f2 | 11.0 us | 500W | dB | 2/17/05 |
| TOCSY | dB | f2 | us | 500W | dB | |
| Dec | dB | f2 | 54.0 us | 500W | dB | |
| LowDec 1 | +11.0 dB | f2 | 80.0 us | 500W | dB | 2/17/05 |
| LowDec 2 | dB | f2 | 160.0 us | 500W | dB | |
| | | | | | | . |
15N | hard | -5.9 dB | f3 | 40.0 us | 300W | | 2/17/05 |
| | -5.2 dB | f3 | 45.0 us | 300W | dB | 11/23/05 |
| | +4.0 dB | f3 | 150.0 us | 300W | dB | 2/17/05 |
| | +7.0 dB | f3 | 200.0 us | 300W | dB | 11/23/05 |
| | | | | | | . |
2H | | -1.5 dB | f4 (lockswitch) | 90.0 us | 300W | | 2/17/05 |
| | +8.75 dB | f4 (lockswitch) | 250 us | 300W | | 2/17/05 |
3. Soft Pulses:
Nucleus | Type | Channel | Pulse | Attenuation | Amplifier | Calculated | Date |
---|
13C | G4.256 340us | f2 | 90 deg | dB | 500W | | |
| G4.256 680us | f2 | 90 deg | dB | 500W | | |
| G3.256 200us | f2 | 180 deg | dB | 500W | | |
| G3.256 640us | f2 | 180 deg | dB | 500W | | |
| G4tr.256 340us | f2 | 90 deg | dB | 500W | | |
| Sinc1 160us | f2 | 90 deg | dB | 500W | | |
| Sinc1 160us | f2 | 180 deg | dB | 500W | | |
1H | Sinc1.1000 1.2ms | f1 | 180 deg | dB | 100W | | |
G3 and G4 pulses were determined with spoal=0.5, on-resonance
4. Miscellaneous:
Magic angle on GREAT III: | factor Z/XY = N/A |
Offset Ca - carbonyl (56 - 175ppm): | -18109Hz |
1H offset @ 4.7 pppm: | 2820 Hz |
Main Field: | 800.19 MHz |
5. Conditions for pulse determination:
- Samples:
- 1H-pulses: 2mM Sucrose + 0.5mM DSS + 2mM NaN3, 90% H2O 10% D2O
- X-pulses: 13C, 15N labled Urea in DMSO
- 13C-shift: 48.5ppm, J=138.5Hz
- 15N-shift: 76.0ppm, J=88.5Hz
- Temperature: 298K
- RF-Filters:
- 1H = none
- X = Highpass 800MHz 13C-Pass; 2H-Stop
- Y = Lowpass 800MHz 15N-pass; 2H-Stop
- Z = none
- Attenuators at amplifier input: none
- 1H = 4dB
- X = 4dB
- Y = 0dB
- Z = 0dB
- Cortab combinations done 3-27-2002 (mc) XWIN-NMR 3.0:
- 1H: --> FCU 1+2, Mult, Mod, BLA, redone after amplifier change 1-13-2004
- 13C:
- FCU 1+2, 500_1: Mult, Mod, BLA
- FCU 3+4, 300_5: Mult, Mod, BLA, 300_6, 500_1: BLA
- 15N:
- FCU 1+2, 500_1: Mult, Mod, BLA
- FCU 3, 300_5: Mult, Mod, BLA, 300_6, 500_1: BLA
- Modify the file: /u/exp/stan/nmr/lists/cpd.hwt/cortab :
- At the end of the file add the line "jump to 1", add 4ms delay.
- Do this only for Amplifier Cortab!
|