603: Cryoprobe 5mm TCI-Z, #Z44896_0077


In principle: Do NOT trust these values blindly - verify them if questionable!

Amplifiers can go bad or your sample environment like salt, temperature, ... might alter the pulse width!

IMPORTANT POWER REFERENCES FOR CRYOPROBES (full document)

4.2 Pulse and power recommendations

 

Note: All values require a minimum repetition period d1 +aq of 1sec 2)

Important: observe the separate notes on the foIIowing pages!

Important: Be CAREFUL when pushing these limits!

 

 

5mm 600 MHz TXI CryoProbe TM

1H

 

Hard pulse 3)

(max. length 360o)

8 usec

Hard pulse for lossy sample

Power level corresponding to 8.0 usec pulse for non-lossy samples

trim pulse p28 4)

2 msec @ 10 usec

TOCSY spin lock

200 msec @ 25 usec

400 msec @ 35 usec

ROESY spin lock

Up to CW for power level

corresponding to a 100 usec pulse

WALTZ 16 decoupling

during 13C-detection

Up to CW for power level corresponding to a 100 usec pulse

DIPSI2-decoupling in triple resonance

400 msec @ 50 usec

13C

 

hard pulse 6)

(max. length 360o)

15.0 usec

trim pulse 5)

2msec @ 25 usec

CC spin lock 5)

20msec @ 25 usec

GARP decoupling 7)

140msec @ 55 usec

selective pulses 8)

G4: 320 usec

Q3: 256 usec

CHIRP: 2ms @ 25 usec

15N

 

hard pulse 6)

(max. length 360o)

40.0 usec

GARP4 decoupling 7)

140 msec @ 170 usec

CPMG T2 9)

250 msec @ 80 usec (see warning 9))

2H

 

hard pulse

(max. length 360o)

150 usec

WALTZ16 decoupling

100 msec @ 250 usec

Z-Gradient

 

Absolute max. current 10)

10 A

Max. overall length 11)

10 ms @ 10A

 

Date: 25.10.2004


1. General Information:

  • Shimfile: chloshim.600 is a good starting point for aqueous samples
  • Read Pulses from Master List:
    • read in pulse sequence and Parameter file
    • Type getprosol or click on the test tube symbol in ased
    • For proton pulses type: getprosol 1H , where p1 is the determined pulse length [us] at the power level pl1 [dB].

2. Hard Pulses:

Nucleus Type Attenuation Channel 90deg Pulse Amplifier Calc. Date
1H 3.8 dB f1 8.40 us 100W 03/03/09
waltz16 f1 70.0 us 100W
TOCSY f1 25 us 100W
ROESY f1 125.0 us 100W
.
13C hard -2.05 dB f2 15.0 us 300W 03/03/09
AcGly TOCSY +2.30 dB f2 25.0 us 300W dB 03/03/09
TOCSY +4.45 dB f2 32.0 us 300W dB 03/03/09
Dec +10.70 dB f2 65.0 us 300W dB 03/03/09
LowDec 1 +12.55 dB f2 80.0 us 300W dB 03/03/09
.
15N hard -0.50 dB f3 40.0 us 300W 03/03/09
+6.60 dB f3 90.0 us 300W dB 03/03/09
+13.60 dB f3 200.0 us 300W dB 03/03/09
.
2H dB 2HTX 150 us 20W
AcGly dB 2HTX 300 us 20W dB

3. Soft Pulses:

Nucleus Type Channel Pulse Attenuation Amplifier Resonance Date
13C , Cali/CO Q5.1000 320us f2 90 deg -0.50 dB300W on/off 10/17/08
Cali/CO Q5tr.1000 320us f2 90 deg -0.90 dB300W on/off 10/17/08
Cali/CO Q5tr.1000 410us f2 90 deg +1.20 dB300W on 10/17/08
Cali/CO Q5tr.1000 410us f2 90 deg +1.25 dB300W off (-132ppm) 10/17/08
Q5.1000 1500us f2 90 deg +13.0 dB300W on/off 10/17/08
Cali/CO Q3.1000 256us f2 180 deg +0.70 dB300W on 10/17/08
Ca/CO Q3.1000 410us f2 180 deg +4.80 dB300W on 10/17/08
CO dec Q3.1000 768us f2 180 deg +10.30 dB300W on 10/17/08
Ca Q3.1000 1000us f2 180 deg 12.75 dB300W on 10/17/08

4. Miscellaneous:

Main Field: 600.13 MHz
BSMS Field: -1093 units (8/10/2007)

5. Conditions for pulse determination:

  • Sample: 13C, 15N labled Urea in DMSO
    13C-shift: 48.5ppm, J=138.5Hz
    15N-shift: 76.0ppm, J=88.5Hz
    Acetyl Glycine 13C/15N, in D2O
    X-pulses: 13C, 15N labled Acetyl Glycine
    13C-shift: 22.1ppm, J=128.6Hz (1H 1.96 ppm )
    15N-shift: 76.0ppm, J=88.5Hz
  • Temperature: 298K
  • RF-Filters: none
  • Attenuators at amplifier input: NONE