501: 2.5mm TXI-Z, #Z8341-0011
In principle: Do NOT trust these values blindly - verify them if questionable!
Amplifiers can go bad or your sample environment like salt, temperature, ... might alter the pulse width!
1. General Information:
- Shimfile: 8341_0011ws.shim is a good starting point for aqueous samples
2. Hard Pulses:
Nucleus | Type | Attenuation | Channel | 90deg Pulse | Amplifier | Calc. | Date |
---|
1H | | 3.78 dB | f1 | 8.0 us | 100W | | 11-28-03 |
| waltz16 | 23.60 dB | f1 | 80.0 us | 100W | 23.78 dB | 12-24-03 |
| TOCSY | | f1 | 25-35.0 us | 100W | | |
| ROESY | 25.30 dB | f1 | 100.0 us | 100W | 25.72 dB | 12-24-03 |
| | | | | . | | |
13C | hard | -1.20 dB | f1 | 15.0 us | 300W | | 12-24-03 |
| LowDec 1 | +13.30 dB | f1 | 80.0 us | 300W | 13.34 dB | 12-24-03 |
13C | hard | -1.25 dB | f2 | 15.0 us | 300W | | 12-24-03 |
| TOCSY | +5.20 dB | f2 | 32.0 us | 300W | 5.33 dB | 12-24-03 |
| Dec | dB | f2 | 54.0 us | 300W | dB | |
| LowDec 1 | +12.90 dB | f2 | 80.0 us | 300W | 13.29 dB | 12-24-03 |
| LowDec 2 | dB | f2 | 160.0 us | 300W | dB | |
| | | | | . | | |
15N | hard | | f3 | 40.0 us | 300W | |
| | +0.2 dB | f3 | 45.0 us | 300W | | 12-24-03 |
| | | f3 | 50.0 us | 300W | | |
| | +13.16 dB | f3 | 200.0 us | 300W | +13.16 dB | 12-24-03 |
| | | | | . | | |
2H | | dB | 2HTX | us | 20W | | |
| | dB | 2HTX | us | 20W | | |
| | dB | 2HTX | 300 us | 20W | | |
3. Soft Pulses:
Nucleus | Type | Channel | Pulse | Attenuation | Amplifier | Calc. | Date |
---|
13C | G4 400us | f2 | 90 deg | - dB | 300W | | |
| G4 640us | f2 | 90 deg | + dB | 300W | | |
| G3 256us | f2 | 180 deg | - dB | 300W | | |
| G3 768us | f2 | 180 deg | + dB | 300W | | |
| G4tr 409us | f2 | 90 deg | - dB | 300W | | |
| Sinc1 160us | f2 | 90 deg | + dB | 300W | | |
| Sinc1 160us | f2 | 180 deg | + dB | 300W | | |
1H | sinc1.1000 1.2ms | f1 | 180 deg | + dB | 100W | | |
4. Miscellaneous:
Offset Ca - carbonyl (56 - 175ppm): | Hz |
1H offset @ 4.7 pppm: | 2351 Hz |
Main Field: | 500.13 MHz |
BSMS Field: | +8369 units (12-24-2003) |
5. Conditions for pulse determination:
- Sample:
- 1H-pulses: 2mM Sucrose + 2mM NaN3, 90% H2O 10% D2O
- X-pulses: 13C, 15N labled Urea in DMSO
- 13C-shift: 48.5ppm, J=138.5Hz
- only carbonyl is labled, o1p = 160.58ppm, f1 direct, f2 with zgf2
- 15N-shift: 76.0ppm, J=88.5Hz
- Temperature: 298K
- RF-Filters:
- 1H = 1H Bandpass, 2H Stop
- X = 0 - 31P Lowpass Filer, rejects 1H
- Y = 0 - 15N Lowpass Filer, 2H Stop
- 2H = 2H Pass, 15N Stop
- Attenuators at amplifier input: none
- 1H = 0dB
- X = 0dB
- Y = 0dB
- Cortab combinations done 3-7-2002 XWIN-NMR 3.0:
- 1H --> FCU 1+2, BLA
- 13C --> FCU 1+2, X-Amplifier: BLA
- 15N --> FCU 3, Y-Amplifier: BLA
- 31P --> FCU 1+2, X-Amplifier: BLA
|